Sysmex XP-300 14-31
Technical Information
Revised October 2012
6) NEUT% [W-LCR (WBC-Large Cell Ratio)]
Ratio of neutrophils to whole WBC
2. WBC research parameters
Each parameter shows the following particle counts contained in 1 µL of whole blood.
• ResearchW: Particle counts between LD and UD of the WBC histogram
• ResearchS: Particle counts between LD and T1 of the WBC histogram
• ResearchM: Particle counts between T1 and T2 of the WBC histogram
• ResearchL: Particle counts between T2 and UD of the WBC histogram
• Analysis result display screen
Analysis results are displayed on the fourth analysis result screen and fourth stored data detail screen.
• Print format
ResearchS, ResearchM, ResearchL: Can be printed in Type1 or Type2.
ResearchW: Can be printed in Type1, Type2 or Type3.
• Interface protocol
ResearchS, ResearchM, ResearchL: Can be output to the host computer in XP format.
ResearchW: Can be output to the host computer in XP format or pocH format.
For details, contact your local Sysmex service representative.
WBC research parameters are for research purposes only.
Do not use analysis results for WBC research parameters in patient diagnosis.