Sysmex XP-300 14-33
Technical Information
Revised October 2012
• WBC and calculation parameters versus analysis value error flags
1) 1A Histogram having high LD frequency with troughs T1 and T2
WL flag is added to all WBC parameters (WBC, LYM%, MXD%, NEUT%, LYM#, MXD#,
NEUT#).
• WBC Histogram Error-WL (1A)
WBC Histogram Error Flag Output
[1] [2]
Case
No.
LD T1 T2 UD WBC
LYM%
MXD%
NEUT%
LYM#
MXD#
NEUT#
HOST
1 High WLWLWLWLWLWLWL 1
2A × T1 T1 T1 T1 T1 T1 5
2B × HighWUT1T1T1T1T1T1 5
3A High × F1 T2 T2 F1 T2 T2 6
3B High×HighWUF1T2T2F1T2T2 6
3C × T2 T2 T2 T2 6
3D ×HighWU T2 T2 T2 T2 6
4A High F1 F2 F1 F2 7
4B High High WU F1 F2 F1 F2 7
4C High High F1 F2 F3 F1 F2 F3 7
4D High High High WU F1 F2 F3 F1 F2 F3 7
5A High F2 F3 F2 F3 8
5B High High WU F2 F3 F2 F3 8
6 High WU 2
7 (The particle count equal to or
less than LD is higher than the
range)
PLT parameter is flagged with AG. A
[1]: Output flag to LCD screen, printer.
[2]: Output flag to host computer.
: Analysis result of each discriminator is normal.
High: Frequency for each discriminator is higher than the range.
×: A trough is not clear and cannot be detected.
: Any of “,” “×,” and “High” is applicable.
WBC
UDLD T1 T2
300200100 [fL]
WBC WL 8.5 [×10
3
/µL]
LYM% WL 39.3 [%]
MXD% WL 17.6 [%]
NEUT% WL 43.1 [%]
LYM# WL 3.4 [×10
3
/µL]
MXD# WL 1.5 [×10
3
/µL]
NEUT# WL 3.6 [×10
3
/µL]