EasyManua.ls Logo

SYSMEX XP Series - Page 264

SYSMEX XP Series
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14-34 Sysmex XP-300
Technical Information
Revised October 2012
2) 1B Histogram with high LD and trough T1 but without T2
WL flag is added to all analysis results for WBC, LYM% and LYM#. WL flag is added to mixed
cell and neutrophil parameters (MXD%, NEUT%, MXD#, NEUT#) and their data are not
output.
([---.-])
WBC Histogram Error-WL (1B)
3) 1C Histogram with high LD but without T1
WL flag is added to WBC and other parameters are not output. Note [+] on WBC numeric data
in the example below. This WBC value exceeds the upper Patient Mark Limits mentioned
earlier.
WBC Histogram Error-WL (1C)
WBC
UDLD T1
300200100 [fL]
WBC WL 2.8 [×10
3
/µL]
LYM% WL 45.9 [%]
MXD% WL ---.- [%]
NEUT% WL ---.- [%]
LYM# WL 1.3 [×10
3
/µL]
MXD# WL ---.- [×10
3
/µL]
NEUT# WL ---.- [×10
3
/µL]
WBC
UDLD
300200100 [fL]
WBC WL +16.4 [×10
3
/µL]
LYM% WL ---.- [%]
MXD% WL ---.- [%]
NEUT% WL ---.- [%]
LYM# WL ---.- [×10
3
/µL]
MXD# WL ---.- [×10
3
/µL]
NEUT# WL ---.- [×10
3
/µL]

Table of Contents