14-34 Sysmex XP-300
Technical Information
Revised October 2012
2) 1B Histogram with high LD and trough T1 but without T2
WL flag is added to all analysis results for WBC, LYM% and LYM#. WL flag is added to mixed
cell and neutrophil parameters (MXD%, NEUT%, MXD#, NEUT#) and their data are not
output.
([---.-])
• WBC Histogram Error-WL (1B)
3) 1C Histogram with high LD but without T1
WL flag is added to WBC and other parameters are not output. Note [+] on WBC numeric data
in the example below. This WBC value exceeds the upper Patient Mark Limits mentioned
earlier.
• WBC Histogram Error-WL (1C)
WBC WL 2.8 [×10
3
/µL]
LYM% WL 45.9 [%]
MXD% WL ---.- [%]
NEUT% WL ---.- [%]
LYM# WL 1.3 [×10
3
/µL]
MXD# WL ---.- [×10
3
/µL]
NEUT# WL ---.- [×10
3
/µL]
WBC WL +16.4 [×10
3
/µL]
LYM% WL ---.- [%]
MXD% WL ---.- [%]
NEUT% WL ---.- [%]
LYM# WL ---.- [×10
3
/µL]
MXD# WL ---.- [×10
3
/µL]
NEUT# WL ---.- [×10
3
/µL]