Sysmex XP-300 14-35
Technical Information
Revised October 2012
4) 2A Histogram without T1
Although the histogram error flag is not added to WBC, all other parameters are flagged with T1
and their data are not output. Note that WBC in the graph exceeds the upper Patient Mark
Limits.
WBC Histogram Error-T1 (2A)
5) 2B Histogram having high UD but without T1
WBC is flagged with WU. All other parameters are flagged with T1 and their data are not
output.
6) 3A Histogram with high T1 but without T2
WBC is not flagged. F1 flag is added to lymphocyte parameters (LYM%, LYM#). Mixed cell
and neutrophil parameters (MXD%, NEUT%, MXD#, NEUT#) are flagged with T2 and their
data are not output.
WBC Histogram Error-T2 (3A)
7) 3B Histogram with high T1 but without T2, and high in UD frequency
WBC is flagged with WU and lymphocyte parameters (LYM%, LYM#) are flagged with F1.
Mixed cell and neutrophil parameters (MXD%, NEUT%, MXD#, NEUT#) are flagged with T2
and their data are not output.
WBC
UDLD
300200100 [fL]
WBC +15.4 [×10
3
/µL]
LYM% T1 ---.- [%]
MXD% T1 ---.- [%]
NEUT% T1 ---.- [%]
LYM# T1 ---.- [×10
3
/µL]
MXD# T1 ---.- [×10
3
/µL]
NEUT# T1 ---.- [×10
3
/µL]
WBC
UDLD T1
300200100 [fL]
WBC 5.2 [×10
3
/µL]
LYM% F1 29.7 [%]
MXD% T2 ---.- [%]
NEUT% T2 ---.- [%]
LYM# F1 1.5 [×10
3
/µL]
MXD# T2 ---.- [×10
3
/µL]
NEUT# T2 ---.- [×10
3
/µL]