Uster Technologies, Inc. Chapter 7
Testing Procedures • Page 7-8
Note:
Allow the balance time to stabilize and make the correct
reading before removing the fiber. The system will not accept the results
until a stable reading is taken.
When the sample weight is within range, the weight is displayed in
the WGT field on the
HVI
SPECTRUM
TESTING screen.
Insert the weighed sample into the micronaire chamber and close the
lid. The micronaire test will begin automatically.
The chamber lid will open once the test is completed and the sample
is ejected from the chamber.
The micronaire value will be displayed in the box for MIC in the
Micronaire
area.
The maturity of the cotton will be displayed in the box for MAT after
the length/strength measurement tests are complete.
Note:
Maturity Index is an optional setting.
If the Micronaire value is outside the set limits:
Select the NEW TEST button to accept the value and proceed with
testing.
To reject the value and perform another test sequence on the sample,
press the REJECT TEST button and then press the NEW TEST button.
Checking the “…AND NEP” box allows you to reject NEP data as
well as other data.
Note: If an error message appears repeatedly, it could indicate an air
leakage in the micronaire chamber. In such a case, perform a
micronaire calibration check to ensure that it is accurate. If necessary,
contact an Uster Technologies’
Service technician for further support.