USTER HVI SPECTRUM - Calibration Test

USTER HVI SPECTRUM
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USTER
®
HVI SPECTRUM
Instruction Manual
Uster Technologies, Inc. Chapter 5
System Check • Page 5-5
2.1.2 Calibration Test
To begin a calibration test:
1. Enter the Lot ID and Bale ID information in the corresponding
fields.
2. Begin the testing procedure by weighing and measuring the
sample. (R
efer to Chapter 4 of this manual for complete
instructions.)
3. Press the START TEST button on the SYSTEM CHECK MICRONAIRE
screen.
4. The micronaire values will be displayed on the screen as the
sample is tested.
o The mean will be displayed on the screen after 2 or more tests
are completed.
o The StdDev and %CV will be displayed after 3 or more tests
are completed.
5. Once testing is complete, press the END TEST button.
Note:
Reporting is not enabled until all testing has ended and the
END
TEST
button has been selected.
6. The QUICK REPORTS button enables you to print a report. Refer to
Chapter 9, Quick Reports, for additional information regarding this
function.