3.4 Trash Calibration
The trash module used in the
HVI SPECTRUM
instrument has been
designed to operate for an extended period of time without requiring
calibration. It is recommended that a
routine
check
be performed
rather than a
routine
calibration
. A calibration should only be done in
cases of extreme swings in temperature or in a highly dusty
environment.
In the event that it becomes necessary to calibrate the trash, follow
these instructions:
1. Select the S
ETUP button from the
HVI
SPECTRUM
MAIN MENU.
2. Select the CALIBRATE button.
3. Choose the TRASH button from the CALIBRATION screen.
Figure 4-11: Calibration Screen
Uster Technologies, Inc. Chapter 4
Calibration • Page 4-17