2.5 Balance Configuration
The
HVI SPECTRUM’s
Micronaire Module is provided with a Mettler
Balance. Please refer to the manufacturer’s User’s Guide that is also
provided with your instrument if you have specific questions on the
balance.
The following is the procedure Uster Technologies uses to configure the
Mettler PB602-S balance. This configuration cannot be changed
without altering the balance hardware. A certified Uster Technologies’
Service technician MUST complete any changes made to the
configuration. For this reason, the following list is provided for
information purposes only.
1. Press hold Cal/Menu button until “Menu” appears.
2. Release button. “Reset” should now be displayed.
3. Press and release Cal/Menu button.
4. Press the F button until “F-nonE” appears.
5. Press and release Cal/Menu button.
6. Press the F button until “Std” appears.
7. Press and release Cal/Menu button.
8. Press the F button until “Unit 1 g” appears.
9. Press and release Cal/Menu button.
10. Press the F button until “Unit 2 g” appears.
11. Press and release Cal/Menu button.
12. Press the F button until “A.2Ero” appears.
13. Press and release Cal/Menu button.