Uster Technologies, Inc. Chapter 4
Calibration • Page 4-3
If a "Fail" condition occurs during a calibration, changes are made to
the parameters. However, ONLY those parameters that are critical
for calibration to be brought to a "Pass" condition will change. In
other words, no changes are made to those parameters that pass
calibration originally.
For example: Assume that length and uniformity pass the
original calibration, but there is a "Fail" condition for strength
(perhaps as a result of a change in the relative humidity). During
the next calibration procedure, the slope and offset for strength are
all that would be changed. The slopes and offsets for length and
uniformity would remain at their original values.