C H A P T E R 7 – T E S T S A N D T E S T C O N D I T I O N S
AT5600 User Manual 98-119 issue 14 Page 159
Choosing test conditions
The recommended test conditions depend on the inductance of the energized
winding; they are given in the table below assuming that the energized winding is
the one with the highest number of turns:
Note that a ‘fixed’ voltage signal cannot be used with an ‘Auto’ frequency.
Please also see
14.6.35 for specific advice on programming the test, and
10.2.10 for test accuracy specification
Best Practice for Centre Taps / AutoTransformers
Please also see the section in 7.13 TURNS RATIO for best practice when
performing LVOC on Centre Taps / AutoTransformers