C H A P T E R 10 S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14 Page 219
10.2.11. IR Test
The accuracy of the test voltage is 3% +/-50V.
The accuracy is based on the chosen test conditions, and is given by:
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
C
is the calibration accuracy (%)
A
I
is the correction for current level (%)
A
Z
is the correction for impedance (%)
where:


 󰇛

󰇜

󰇛

󰇜
where:
I
M
is the measured current (A)
Z
M
is the measured IR value (M)