C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S
AT5600 User Manual 98-119 issue 14 Page 160
7.40. ILK Leakage Current Test
Where Used
The leakage current test is needed for transformers used in certain applications,
mainly medical, where it is an additional safety requirement.
Measurement Conditions
The test is typically carried out by applying the test voltage between all the
primary terminals shorted together, and all the secondary terminals and screen
shorted together.
The voltage and frequency applied are normally the operating voltage and
frequency of the transformer.
Please also see
14.6.36 for specific advice on programming the test, and
10.2.26 for test accuracy specification