C H A P T E R 1 4 - A T E D I T OR
AT5600 User Manual 98-119 issue 14 Page 309
14.5 .12.3 Test Group ing
Tests can be programmed in any order you wish, and the results for each individual test
will be independent of where they are in the program.
However, the AT5600 has 6 different power sources, and to get the best speed
execution time we recommend that tests are grouped using the following table.
This minimizes change over time between the different sources.
Please note that the table and “Best Speed Grouping” column does not indicate any
overall ORDER for the tests, just the best grouping.
Tests in the same group can be run in any order, and the groups can be run in any order
(e.g. most customers always start with R to check windings)
For example
LS RLP Z HPAC
would execute faster than
LS HPAC RLP Z
as HPAC is in a different grouping from the other 3
Or
R LS RLS MAGI VOC
would execute faster than
R LS MAGI RLS VOC
as MAGI and VOC are in the same group.