C H A P T E R 1 4 - A T E D I T OR
AT5600 User Manual 98-119 issue 14 Page 314
14.6.3. RLS or RLP - Equivalent Series or Parallel
Resistance
The equivalent series or parallel resistance is obtained from a measurement of
the complex impedance. An AC voltage is applied across the selected winding.
The voltage across and current through the winding are then measured using
harmonic analysis. The measured voltage is divided by the current to obtain a
complex impedance and the equivalent series or parallel resistance calculated.
Select eitherRLS Equivalent Series Resistance’ orRLP Equivalent Parallel
Resistance’ from the Available Tests window, to display:
or: