C H A P T E R 1 4 - A T E D I T OR
AT5600 User Manual 98-119 issue 14 Page 287
CANCEL Terminate this program run, and return to the Execute display, from
where you can start the program again for the next transformer in the batch.
Date and Time on Reports
AT3600+ ATI only
Enable this option to have the date and time on the printer results printout.
AT5600
All printed results using the USB printer will have data and time on them
regardless of this setting.
Disable test failu res due to ALC error
Under test conditions where the load is mismatched to the requested test signal,
then the voltage or current may not be met (to within nominal spec of 1%).
In these cases, the unit will return a STATUS ERROR of “0020" (see section
9.1.3).
This indicates that, while the measured parameter may be within your requested
limits, the test itself is a FAIL, as the requested test conditions were not met.
This is analogous to ALC (Auto Level Control) on most popular LCR meters.
With this option enabled, the AT5600 will make the PASS/FAIL judgement purely
based on the measured parameter and ignore the “0020” error code.
The AT3600 and ATi units always make the PASS/FAIL judgement on limits
alone, so enabling this option will help users with compatibility to these platforms.
Please note; The AT3600 and ATi firmware do NOT support this option;
Regardless of the option setting, the 0020 status will never be part of the
PASS/FAIL criteria for these units.