C H A P T E R 10 S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14 Page 210
10.2.2. LS, LP, RLS, RLP, LL and C Tests
The accuracy is based on the chosen test conditions, and is given by:
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
C
is the calibration accuracy (%)
A
V
is the correction for voltage level (%)
A
I
is the correction for current level (%)
A
Q
is the correction for Q factor (%)
where:


󰇛

󰇜






󰇛󰇜
 󰇛
󰇜
󰇛󰇜
where:
F
M
is the test frequency (Hz)
V
M
is the measured voltage (V)
I
M
is the measured current (A)
Q
M
is the measured Q factor