C H A P T E R 10 S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14 Page 243
In the standard set-up, the three status outputs are active Lo signals which signal
Busy, Pass and Fail. They could, for example, be connected as status inputs to a
robot controller, a bin selector, or indicator large lamps to show the status.
All the outputs are fitted with 1k protection resistors, and therefore they cannot
directly drive external high current loads such as indicators, or bin-select relays.
Beep Output (Pin 1)
This is an active Hi output which copies the internal beeper.