C H A P T E R 1 4 - A T E D I T OR
AT5600 User Manual 98-119 issue 14 Page 368
To program the test:
1. Enter the signal, frequency and bias current required for the test.
2. Select (mouse click the button) the integration time you require.
‘Medium’ is the default setting. ‘Long’ will give you more stable readings
(for tighter limits) at the expense of test time, and ‘Short’ will test at the
maximum speed, but may give slightly noisier readings. See 7.3
Explanation of Integration.
3. Enter the terminal names between which the inductance will be measured.
By pulling down the menu alongside the fill-in box, you may select the high
and low terminals for each test.
4. Select (mouse click the button) the type of limits you require, and enter the
values.
% Enter a nominal value and then the limits as negative
and positive values.
>< Enter minimum and maximum values.
> Enter just a minimum value.
< Enter just a maximum value.
The Measure button may be used here as outlined in section 14.5.9.
5. Select OK. The test specification will then be displayed in the program
window.