EasyManua.ls Logo

Voltech AT5600 - Page 434

Voltech AT5600
438 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C H A P T E R 1 6 - C H A N G E L O G
AT5600 User Manual 98-119 issue 14 Page 434
V 0.007.246 23 APR 2018
Functionality
Implemented filter control in range offsets.
Added a support for a hardware-based watchdog to recover from severe disruption.
Editor results return now waits for the enumeration to finish. This fixes the comms
drop after a sparkover.
Fixed an issue with high frequency class D measurements reporting 0.0000 due to a
timeout.
Tests
IR - Unified the range configuration in DC remainder maintenance to I prove IR test
RLOS - Fixed limits for measure mode for RLOS.
RLOS - Fixed parsing of RLOS measure mode command.
L2 - Trimming is based on the measured ratio rather than the nominal ratio.
L2 - Fixed L2 returning a trimming error on auto secondary signal level.
STRW Fixed bug causing occasional 0.000 results
ILK - Resolved a spurious excessive load error during class D ramp up
TRL - Resolved negative readings for turns ratio by inductance when negative
polarity is measured with no polarity check specified.

Table of Contents

Related product manuals