6166 DC Voltage Current Source Operation Manual
5.4.4 Function Change Timing
5-11
5.4.4 Function Change Timing
Tfun
VS IS 45 ms
VS DIV 45 ms
IS VS 65 ms
IS DIV 65 ms
DIV VS 100 ms
DIV IS 100 ms
Function before change
Function after change
Settling time
Function command
or
key operation