6166 DC Voltage Current Source Operation Manual
8.2 Calibration Method
8-3
8.2 Calibration Method
The 6166 can be calibrated manually or using the GPIB or USB remote commands.
8.2.1 Calibration Points and Tolerance Ranges
For calibration, use the measurement instruments satisfying the required accuracy described in Section
8.1.1, "Measuring Instruments Necessary for Calibration", meeting the tolerance ranges shown in the fol-
lowing table.
Table 8-2 Calibration Tolerance Ranges
Voltage source calibration Current source calibration
Range Calibration point
Resoluti
on
Tolerance
range
[digit]
Range Calibration point
Resolut
ion
Tolerance
range
[digit]
10 mV
+00.00000 mV 10 nV 6
1 mA
+0.000000 mA 10 nA 1
+10.00000 mV 8 +1.000000 mA 1
-10.00000 mV 8 -1.000000 mA 1
100 mV
+000.0000 mV 100 nV 2
10 mA
+00.00000 mA 100 nA 1
+100.0000 mV 3 +10.00000 mA 1
-100.0000 mV 3 -10.00000 mA 1
1000 mV
+0000.000 mV 1 V 1
100 mA
+000.0000 mA 1 A 1
+1000.000 mV 1 +100.0000 mA 1
-1000.000 mV 1 -100.0000 mA 1
1 V
+0.000000 V 1 V 1
+1.000000 V 1
Voltage limiter calibration
-1.000000 V 1 Calibra-
tion point
Resolution
Tolerance range
[digit]
10 V
+00.00000 V 10 V 0.5
+10.00000 V ±1 +0010 V 1 V 0.2
-10.00000 V 1 +1000 V
100 V
+000.0000 V 100 V 0.5
+100.0000 V 1
Current limiter calibration
-100.0000 V 1
Calibration
point
Resolution
Tolerance range
[digit]
1000 V
+0000.000 V 1 mV 0.5
+1000.000 V 1 +001 mA 1 mA 0.02
-1000.000 V 1 +100 mA
The 10 mV to 1000 mV ranges are divided output.