6166 DC Voltage Current Source Operation Manual
6.6 Remote Commands
6-28
6.6 Remote Commands
6.6.1 Command Syntax
The command syntax is defined by the following format.
1. Header
There are two types of headers: common command headers and simple headers. The common command
header has an asterisk mark (*) placed in front of the mnemonic.
The simple header is a functionally independent command that has no hierarchical structure.
Placing a question mark (?) right after the header forms a query command.
2. Space (blank character)
One or more spaces can be used (Spaces may be omitted).
3. Data
When a command requires multiple data sets, data sets are separated by comma (,). A space may be used
directly before or after comma (,). For more information on the data types, refer to Section 6.6.2, "Data
Format."
4. Describing Multiple Commands
The 6166 allows multiple commands to be described consecutively or separated by semicolon (;), com-
ma (,), or space ( ) in one line.
Header
Space (blank
character)
Data
Space (blank
character)
Space (blank
character)