EasyManuals Logo

ADCMT 6166 User Manual

ADCMT 6166
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
6166 DC Voltage Current Source Operation Manual
3.8.5 Setting Program Memory Scan Operations
3-45
Press . The memory recall screen is switched to the program screen. Memory recall oper-
ations are complete.
Example display Memory recall screen
To terminate the program function and return to the source value screen, press or
.
3.8.5 Setting Program Memory Scan Operations
Memory scan is a method to recall the program memory from the specified first number to last number in
order.
There are the following three types of memory scan with different operations.
The following shows the procedure of selecting the scan operations.
Table 3-9 Program Memory Scan Operations
Scan operation Description
Hold Outputs the source values from the specified first number to last number in order
by every trigger input.
The trigger input includes key, remote and external trigger input.
Single Outputs the source values from the specified first number to last number in order at
step time intervals.
Repeat Outputs repeatedly the source values from the specified first number to last num-
ber in order at step time intervals.