Loading...
6166 DC Voltage Current Source Operation Manual
7.3 Test Method
7-3
Table 7-1 Voltage/Current Source/Limiter Check Points
Voltage source
Range Check point VL setting IL setting
+00.00000 mV
10 mV +10.00000 mV
-10.00000 mV
+000.0000 mV
100 mV +100.0000 mV
-100.0000 mV
+0000.000 mV
1000 mV +1000.000 mV
-1000.000 mV
+0.000000 V
1 V +1.000000 V
-1.000000 V
+00.00000 V
10 V +10.00000 V 120 V 10 mA
-10.00000 V
+000.0000 V
100 V +100.0000 V
-100.0000 V
+0000.000 V
1000 V +1000.000 V 1200 V 10 mA
-1000.000 V

Table of Contents