6166 DC Voltage Current Source Operation Manual
7.3 Test Method
7-4
Current source
Range Check point
Standard
resistance
VL setting IL setting
+0.00000 mA
1 mA +1.00000 mA 1 k
-1.00000 mA
+00.0000 mA
10 mA +10.0000 mA 100 10 V 120 mA
-10.0000 mA
+000.000 mA
100 mA +100.000 mA 10
-100.000 mA
Voltage limiter
Check point VS setting IL setting
+0010 V 100 V 10 mA
+0050 V
-0010 V -100 V
-0050 V
Current limiter
Check point Standard resis-
tance
VS setting VL setting
+001 mA 10 10 V 20 V
+100 mA
-001 mA -10 V
-100 mA
Table 7-1 Voltage/Current Source/Limiter Check Points