6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-37
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL
Mem-
ory
Memory setting Nnnn
P
Voltage/current
Nnnn,,,,,P
nnn: memory number 0 to 999
Source range setting
SVRn/SIRn/SDRn
Auto range setting when omitted
Output data setting *1
SOVddd.ddddE+d
SOIddd.ddddE+d
SODddd.ddddE+d
Voltage limit value setting
LMV+d.dddE+d
Default value when omitted
Current limit value setting
LMI+d.ddE-d
Default value when omitted
P: Memory setting end
Thermal electromotive force
Nnnn,,,,,P
Thermocouple selection
TC1 to TC8
Default value when omitted
JIC C1602 version
JSVn
C1602-1995 when omitted
Thermal electromotive force
source value setting *1
TMPdddd.dE+d
Reference function temperature
setting
TMCdddd.dE+d
0ï€ ï‚°C when omitted
P: Memory setting end
Not initialized by the RINI or *RST
command.
N0 -
N?nnn Query of specified memory number *2
Response: Nddd, , , , ,P
or Nddd, , , , ,P
For the specifications of to ,
refer to the above N and P commands or
individual commands.
-
NP? Query of memory setting status
Response: 0 memory setting complete
1 memory setting in progress
0-
*1: The setting range of the source value varies depending on the model, range or function. For more information, refer to Chapter 9,
"SPECIFICATIONS."
*2: The position of the decimal point varies depending on the model, range or function.