6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-36
Source Limit value LMV data Voltage limit value setting
130 V - - -
LMVd.dddE+d
LMV? Query of voltage limit value
Response: LMV+d.dddEd
LMI data Current limit value setting
125 mA - - -
LMId.ddE+d
LMI? Query of current limit value
Response: LMI+d.ddEd
Remote sense RS0 2 Wire
-
RS1 4 Wire
RS? Query of remote sense
Response: RS0 or RS1
Guard switch-
ing
GRD0 INT.GUARD (internal guard)
-
GRD1 EXT.GUARD (external guard)
GRD? Query of guard switching
Response: GRD0 or GRD1
Discharge relay DCR0 OFF
--
DCR1 ON
DCR? Query of discharge relay
Response: DCR0 or DCR1
Polarity type PLM0 Sets the polarity type to bipolar
PLM1 Sets the polarity type to unipolar
PLM? Query of polarity type
Response: PLM0 or PLM1
Count fix at
range change
CFIX0 The count value is not fixed at range
change.
CFIX1 The count value is fixed at range
change.
CFIX? Query of count fix at range change
Response: CFIX0 or CFIX1
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL