6166 DC Voltage Current Source Operation Manual
1. PREFACE
1-1
1. PREFACE
This chapter describes the accessories, operating environment, precautions, and operating check for person-
nel who operate the 6166. Read through this manual before use.
1.1 Product Overview
The 6166 is a DC voltage current source that adopts a PWM system in the reference voltage generation block.
The 6166 is ideal for the working standard capable of manual setting by keyswitches, 1000-channel memory
setting, and remote setting.
In addition, its thermal electromotive force output function allows thermometers to be calibrated easily by
entering temperature values.
• Maximum output: 1199.999 V, 119.9999 mA
• Voltage range: 10 mV to 1000 mV (divider output), 1 V to 1000 V
• Current range: 1 mA to 100 mA
• Display digit: 6
1
/
2
digits
• Resolution: 10 nV (divider output, 10 mV range)
1 V (1 V range)
1 nA (1 mA range)
• Overall accuracy: 25 ppm per 90 days
• Stability: 5 ppm per day
• Voltage limiter: 10 V to 1250 V
• Current limiter: 1 mA to 125 mA
• Thermal electromotive force:
Corresponding to eight types of thermocouples: T (CC), J (IC), E (CRC),
K (CA), S (PR10), R (PR13), B (PR30), N
Compliant with JIS C1602-1995 and JIS C1602-1981
• Compliance voltage: 120 V
1200 V in the 1 mA and 10 mA ranges (OPT6166+20)
• Interface: GPIB and GPIB as standard
BCD (OPT6166+04)
• Alarm detection: Limiter, overload, overheat, fan stop
• External control signal: TRIGGER IN, INTRLOCK
• Compatibility: Compatible remote commands with the 6161 and the TR6120A