6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-38
Mem-
ory
Memory setting RSAV Saves memory data.
--
RLOD Loads memory data.
--
RCLR Initializes memory data.
Data in the non-volatile memory is
initialized.
Saved data is not initialized.
--
Scan First number
Last number
SC First number and last number setting
SCmmm,nnn
mmm: First number 0 to 999
nnn: Last number 0 to 999
0999 - -
SC? Query of first number and last number
Response: SCmmm,nnn
mmm: First number 0 to 999
nnn: Last number 0 to 999
Step time SP Step time setting
SPdd dd: 01 to 99 (sec)
1
--
SP? Query of step time
Response: Spdd
Scan operation SCM0 Hold scan
--
SCM1 Single scan
SCM2 Repeat scan
SCM? Query of scan operation
-
Response: SCM1 to SCM2
Memory num-
ber
RCLMnnn Specifies memory number.
0-
RCLMnnn,: memory number 0 to 999
RCLM? Query of present memory number
-
Response: RCLM000 to RCLM999
Scan start *TRG Starts, pauses or resumes scan.
Sending *TRG during scan operation
pauses it. Sending *TRG again during
pause resumes it.
-
Scan stop SWSP Stops scan.
-
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL