6166 DC Voltage Current Source Operation Manual
6.6.5 6161-Compatible Remote Command List
6-45
Source Voltage limit
value
VL data Voltage limit value setting (1 V step)
130 V 130 V - - -
VLdddd
VL? Query of voltage limit value
Response: VLdddd
L1 Voltage limit value 10 V
---
L2 Voltage limit value 130 V
L3 Voltage limit value 300 V
L4 Voltage limit value 1250 V *6
IL data Current limit value setting (1 mA step)
125 mA 125 mA - - -
ILddd
IL? Query of current limit value
Response: ILdddd
L5 Current limit value 5 mA
--
L6 Current limit value 13 mA
L7 Current limit value 50 mA
L8 Current limit value 125 mA *6
L? Query of voltage or current limit
Response: L1 to L8
Remote sense SEN0 2 Wire
-
SEN1 4 Wire
SEN? Query of remote sense
Response: SEN0 or SEN1
Guard switching GRD0 INT.GUARD (internal guard)
-
GRD1 EXT.GUARD (external guard)
GRD? Query of guard switching
Response: GRD0 or GRD1
*6: Commands compatible with TR6120A
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Scan
stopped
Scan
running