6166 DC Voltage Current Source Operation Manual
A.3.1 GPIB/USB Remote Execution Time (Typical Value)
A-7
2. Source value data read time
Source range
change
SVR4 (Original range: SVR7) Output status: Operate
Other: default
676 673
SVR5 (Original range: SVR4)
170 169
SVR6 (Original range: SVR5)
181 180
SVR7 (Original range: SVR6)
296 294
SIR1 (Original range: SIR3)
387 385
SIR2 (Original range: SIR1)
672 670
SIR3 (Original range: SIR2)
776 775
SDR2 (Original range: SDR4)
170 168
SDR3 (Original range: SDR4)
170 168
SDR4 (Original range: SDR2)
58 58
Voltage source*
SOV<d.ddddddE+d> (No range change) Output status: Operate
Other: default
88
(Range change)
143 to 661 142 to 661
Current source*
SOI<d.ddddddE+d> (No range change)
88
(Range change)
17 to 520 17 to 517
Divider voltage
source*
SOD<d.ddddddE+d>
99
Thermal electro-
motive force
source*
TMP<d.ddddE+d>
48 48
Voltage limit*
LMV<d.dddE+d>
77
Current limit*
LMI<d.dddE+d>
77
* The command with <data> is different in processing time according to the data length.
Item Command Condition
GPIB
Unit [ms]
USB
Unit [ms]
Query command
SOV? Default
33
SOI?
33
TMP?
55
SOD?
33
Item Command Condition
GPIB
Unit [ms]
USB
Unit [ms]