EasyManuals Logo

Agilent Technologies E6198B User Manual

Agilent Technologies E6198B
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #249 background imageLoading...
Page #249 background image
Register Definitions B
E6198B Switch/Load Unit User Manual B-33
• Power on/Reset state is no current sense channels selected
and differential amp disconnected from the Isense bus.
Load Disconnect 1-8 (W) Base + 4
h
This register controls the relays that connect Loadx to Chanx
for channels 1-8. (See Figure B-10.)
• Write Only
• State: 0 = open, 1 = closed
• Power On/Reset State = 0 (all channels open)
Load Disconnect 9-16 (W) Base + 5
h
This register controls the relays that connect Loadx to Chanx
for channels 9-16. (See Figure B-10.)
• Write Only
• State: 0 = open, 1 = closed
• Power On/Reset State = 0 (all channels open)
Power Select 1-8 (W) Base + 6
h
This register controls the Pull Up/Down Power Select armature
relays for power to channels 1-8. (See Figure B-10.) These relays
are Form C and the register uses positive logic: 1 = NO shorted
to COM, NC is open; 0 = NC shorted to COM, NO is open. The
Bit 76543210
Purpose
Chan8 Chan7 Chan6 Chan5 Chan4 Chan3 Chan2 Chan1
Setting state state state state state state state state
Bit 76543210
Purpose
Chan16 Chan15 Chan14 Chan13 Chan12 Chan11 Chan10 Chan9
Setting state state state state state state state state

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6198B and is the answer not in the manual?

Agilent Technologies E6198B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6198B
CategoryTest Equipment
LanguageEnglish

Related product manuals