38477BA2B
2 – 140 25.5.09
2 Installation Volume 2
Explosion hazard!
If the screw thread on the cover or on one of the four screwed
cable glands is damaged, the Uni-Probe cannot be used any more
in ex-protected areas. If the Uni-Probe housing receives a mechan-
ical blow, e.g. because it is dropped, then you have to return the
Uni-Probe to BERTHOLD TECHNOLOGIES for inspection.
Carefully install the mounting brackets and fixtures using the draw-
ings of the shielding and taking into account the circumstances at
the measuring site.
Risk of injury!
The weight of the source shielding may be up to several 100kg,
depending on the version.
Make sure that the mechanical stability of the mounting devices
matches the weight of the shielding. The operating staff must be
wearing hard hats and safety shoes whenever doing any work on
the shielding.
The size and position of the measurement range to be covered are
determined in the projection phase of the measuring site and
defined in drawings, sketches or details in writing. For assembly,
these specifications have to be observed closely, since deviations
may cause malfunctions of the measuring system.