EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - Test Calculation; I;O Test

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 329
Volume 3 Menu Structure
3
2.32 TEST CALCULATION
1 CPS
Enter a count rate and check the level reading. Level shows the
expected value.
The Test Mode is enabled if you enter a count rate >0. During Test
Mode the failure relay indicates an alarm.
IMPORTANT
Reset the test count rate to “0” as soon as the test is over to con-
tinue in the normal measurement mode. Otherwise, only the test
value will be displayed, no measured value.
2 LEVEL
Level reading depending on the calibration and the defined test
count rate.
3 CURRENT
Check if the current output matches your test level reading.
Tip
A test calculation should be performed after every calibration to
ensure that the device has been set up correctly.
2.33 I/O TEST
1 STATUS DIG. INPUT
Check the status of the digital inputs.
2 DIG. OUTPUT TEST
Opens the relay test window.
3 CURRENT OUTPUT
Current outputs test and adjustment of current output.

Table of Contents