2.16 GENERAL DATA
1 DATE
Read or update the date. The current date is written as follows:
MM / DD / YYYY.
The correct date is important for the automatic decay compensa-
tion of the isotope. A battery keeps the clock running even when
power supply is off.
If the Uni-Probe has not been supplied with power for a longer
period of time (several weeks), then the capacitor may be empty. If
the power supply is turned on again, then the clock starts with the
last saved date. Thus, the error due to the missing decay compen-
sation is minor to negligible. In this case, the date has to be
updated.
2 TIME
View and set the time. The current time is written as follows:
HH:MM:SS.
Differences in time have virtually no effect on the correction of the
decrease in activity. Verification of the time can help to check the
equipment functions. A battery keeps the clock running when
power is off.
3 DESCRIPTOR
The DESCRIPTOR identifies the measurement device. It may be
used as needed. The default value is LB 490. A maximum of 16
characters can be entered.
4 MESSAGE
The term Message can be used as needed. The default value is
UNI-PROBE. A maximum of 32 characters can be entered.
5 T
AG
The term Tag (measurement system) may be used as needed. The
default value is LEVEL. A maximum of 8 characters can be entered.
6 SENSOR NO (starting with software revision 12)
Standard input is the Device ID.
7 ASSEMBLY NO (starting with software revision 12)
8 D
ISTRIBUTOR
BERTHOLD TECHNOLOGIES is entered as Distributor. This term is
fixed and cannot be changed by the operator.