Volume 6 7 Error Handling
6
7.4 Reset
The Uni-Probe can be reset as follows:
•Software reset
Via RESET PHYSICAL BLOCK on the DEVICE menu. The Uni-Probe
will be reset and all parameters will be overwritten by standard
values.
• Hardware reset
By pushing the Reset button on the digital board. The housing
must be open! See Volume 2, page 2–189 of the User’s Manual.
The parameters remain unchanged.
•Power supply
Turning the power off/on also leads to a reset of the Uni-Probe.
The parameters remain unchanged.
7.5 Operation Modes during Measurement
Operating mode Measurement
Normal RUN (normal averaging)
Halt Held
Stop Held
TEST Display according to test count rate
READ-IN Arithmetic averaging of calibration count rate