Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 307
Volume 3 Menu Structure
3
2.9 STATUS CONTROL
1 OP MODE
Shows the current status of the operating mode. The following
modes are possible:
• Normal:
Normal measurement mode =
RUN. Measured values on the
display and the current output are refreshed.
• Measurement stopped:
The measurement has been stopped. The measured value dis-
play and the current output are held. The failure relay indicates
an alarm. See also Instrument Response in Case of Error on
page 3–393.
• Measurement halted:
Current output and measured value display are held. The fail-
ure relay remains in the status normal (no alarm). If the cause
for the halt has disappeared, the measurement automatically
switches to RUN.
• Test Mode:
The measured value display and the current output follow the
test count rate.
2 SHOW ERROR LOG
The error log shows the errors that have occurred last in a table. In
addition, the date and time of each error are displayed. Up to 20
errors can be stored. If more than 20 errors have occurred, only
the last 20 errors will be displayed.
3 REVIEW
This menu item includes additional information. The use and mean-
ing of some of the information is up to the operator, whereas others
are determined by the BERTHOLD TECHNOLOGIES.