7 Error Handling Volume 3
7.2 Error Handling Modes
The operation mode Continue or Stop can be set on the ERROR
HANDLING menu (see page 3–315).
Both operation modes behave as follows:
• Continue measurement:
If the error disappears, the measurement automatically goes to
the RUN mode. If the error relay has indicated an alarm status,
it automatically returns to the normal status. The error mes-
sage is no longer signaled at the current output.
• Stop measurement:
If the error disappears, the error relay remains on alarm. The
error message is still signaled at the current output.
The error message has to be reset.