Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
7 – 719
Volume 7 7 Error Handling
7
7.5 Reset
The Uni-Probe can be reset as follows:
1. Factory Reset
Via the fieldbus software. The parameter for the reset is located
in the function block
RESOURCE. Select Defaults in the RESTART
selection field.
The Uni-Probe will be reset and all parameters will be overwrit-
ten by default values.
2. Hardware Reset
By pushing the Reset button on the digital board.
The housing must be open! See Volume 2 of the User’s Manual.
The parameters remain unchanged.
3. Power supply
Turning the supply voltage off/on also leads to a reset of the
Uni-Probe.
The parameters remain unchanged.