Volume 6 6 Explanations
6
6.3 Linear and Exponential
This function is used only for 1-2 point calibrations.
linear The relation between level and count rate is linear.
Figure 6-4 Linear relation
Linear calibration is used in applications involving rod sources and
point detectors
Exponential The relation between level and count rate is an exponential func-
tion.
Figure 6-5 Exponential
This calibration is used in applications involving point sources and
point detectors.
6.3.1 Exponential One-Point Calibration
If the tank cannot be completely filled after operation has started,
then a one-point calibration can be performed. To this end, enter
an absorption path and product density. Then an empty calibration
is performed for which the initial level value is entered as a per-
centage and the corresponding count rate is entered.
100%
0%
CPS
Meas. range
absorption
path