Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 367
Volume 3 5 Functional Processes
3
5 Functional Processes
The following chapter describes the major functional process that
occur when working with the Uni-Probe.
5.1 Plateau Measurement
If the following display does not come up, then select it, starting
from the Start menu, via the following path:
ACCESS TO SETUP FULL ACCESS SERVICE PLATEAU.
Select
PLATEAU MEASURE.
Select YES.
If you want to continue with the standard setting, then you have to
select
NO. Skip the next 4 windows and continue with START PLA-
TEAU MEASUREMENT NOW?
Enter the high voltage where the plateau measurement should
start and confirm the input with
ENTER.
Enter the high voltage where the plateau measurement should
end and confirm the input with ENTER.