38477BA2B
3 – 368 25.5.09
5 Functional Processes Volume 3
Enter the step size of measurement points and confirm the
entry with
ENTER.
The step size determines the number of value pairs. The larger the
step size, the lower the number of value pairs!
Enter the time of how long a measurement point should be
measured or the count rate is to be averaged and confirm the
entry with
ENTER.
Select
YES to start the plateau measurement.
The plateau measurement has been started.
The following information will be displayed:
• Select Act. HV: HV value which is just being measured.
• act. cps: Count rate.
The count rate is being measured, averaged and updated con-
tinuously.
• timer: remaining measurement time for this measurement
point.
Confirm with
OK.
This completes the plateau measurement.