EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - 4 Calibration; 4.1 Preparing Calibration

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 341
Volume 3 4 Calibration
3
4 Calibration
Prerequisites for calibration with
the HART
®
Communicator
The Level Gauge Uni-Probe LB 490 has to be installed and
connected correctly (see Volume 2, sections 2 and 3).
–The HART
®
Communicator has to be connected correctly.
4.1 Preparing Calibration
For correct calibration of the measuring system, you have to ensure
that the detector and basic settings of the Uni-Probe are correct.
Turn the power supply for the Uni-Probe on.
Turn the HART
®
Communicator on.
After power on, the Start menu appears.
Tip
Push the HOME button to return to the Start menu from any level.
First operation of the system:
Set the basic parameters for the Uni-Probe (see page 3–346).
Calibration of a system that has already been taken into opera-
tion:
If you are sure that the detector and basic configuration of the
Uni-Probe is correct, you can start with the calibration of the mea-
suring system immediately as described in section 4.2.
Otherwise proceed as follows:
If a password has been defined for the Uni-Probe, revoke the
password protection (see page 3–369).
Reset the Uni-Probe to the factory settings (see page 3–342)
Set the parameters for the detector HV (see page 3–344).
Set the basic parameters for the Uni-Probe (see page 3–346).

Table of Contents

Related product manuals