4 Calibration Volume 3
4.1.3 Setting the Basic Configuration
This chapter describes how to perform a minimal basic setting to
prepare the Uni-Probe for calibration. The date is updated and the
isotope is set.
If the following display is not visible, push
HOME to go to the
Start menu.
Select ACCESS TO SETUP.
Select
FULL ACCESS*.
*)
FULL ACCESS or CONFIG. & CALIBR.
Select
CONFIGURATION.
Select
GENERAL DATA.