Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 345
Volume 3 4 Calibration
3
Enter the code as shown in the table in Volume 2, section
"1.6 Detector Codes", on page 2–133.
Save the changes with SEND.
Select
HV SETTINGS.
Select
HV SETUP.
Enter 0 to enable the automatic mode.
Save the changes with
SEND.
Wait until the HV (
HV READING) has been set.
This takes…
– for point detectors: about 2 minutes
– for rod detectors: about 15 minutes
Read off the value
HV READING.
Select HV DEFAULT and enter the read-off value there.
Save the changes with SEND.
This completes the detector HV settings.