Volume 7 7 Error Handling
7
7 Error Handling
Errors are indicated by the error relay. Error messages are dis-
played via the fieldbus control software and/or the process control
system and in the parameter XD_ERROR in the function block
TRANSDUCER.
7.1 Device Response in Case of Error
Use the function block RESOURCE and the parameters
ERROR_LOG_ENTRY and ERROR_LOG_DATA for query (see section
5.10 on page 7–699). You find the parameter Quality in the record
PRIMARY_VALUE in LE_LEVEL.
1 The device first switches to the safe state and then has to be reset by turning off/on the power supply so that
the measurement can be started again.
2) Error 14 always leads to error 13! The error log shows both errors.
Error
number
Error message Quality Error relay Error handling
Measurement
continue
Measurement
stop
0 Fatal Error RAM Bad Alarm Stop Stop
1 Fatal Error ROM Bad Alarm Stop Stop
2 Fatal Error EEPROM Bad Alarm Stop Stop
3 Error CPU Clock Bad Alarm Stop Stop
4 Error Power Fail Bad Alarm Stop Stop
5 Error No Pulse Rate Bad Alarm Continue Stop
6 Error Wrong HV Uncertain Alarm Continue Stop
7 Error HV Deviation Uncertain Alarm Continue Stop
8 Warning Change Multiplier Uncertain Alarm Continue Stop
9 Error TEMP out of range Uncertain Alarm Continue Stop
10 Error Analog Frontend Bad Alarm Continue Stop
11 Error Control countrate Uncertain Alarm Continue Stop
12 Error No Control countrate Bad Alarm Continue Stop
13 Error Watchdog Reset Bad Alarm
1
Continue Continue
14 Error Hardware Reset Bad Normal
2
Continue Continue
17 Error RTCbackup / Battery Uncertain Alarm Continue Stop
18 Error RTC error Uncertain Alarm Continue Stop
19 Error RS-485 Communication Bad Alarm Continue Stop
20 Warning cps < Min Uncertain Normal Continue Continue
21 Warning cps > Max Uncertain Normal Continue Continue
22 Warning Rad.Interference Uncertain Normal Continue Continue