Volume 5 9 Error Handling
5
9.5 Operation Modes during Measurement
The operating mode determines, for example, if a measurement is
currently running or not. Use
DEVICE VIEW PARAMETER STA-
TUS/ERROR LOG
to display the operating mode.
9.6 Error Reset
Error messages are displayed in a message window and / or by a
changed device icon (see page 5–525) in SIMATIC PDM.
The error is reset via modal dialog boxes; this means, that the user
has to be informed about the error before he or she clicks the
OK
button.
Problematic operating states of the device will result in a changed
device icon. This status will be reset only after elimination of the
error. Then these errors will be reset automatically.
Operating
mode
Measurement Current output
Normal RUN (normal averaging) Normal
Halt Frozen. The measurement
is stopped by:
–external input HALT
exceeding the count
rate limits of the
parameter Upper Limit
Value or Lower Limit
Value
Holds the last value
Stop Halted (aborted due to
error message)
Fault current
TEST Display according to test
count rate
Test, follows the test pulse
rate.
READ-IN Arithmetic averaging of
calibration count rate
Holds the last value