EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - Set Multi-Detector Mode

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 371
Volume 3 5 Functional Processes
3
5.3 Set Multi-Detector Mode
The multi-detector mode is needed if several detectors are con-
nected in series to cover a larger measuring range.
In the multi-detector mode, one detector is used as Main-Unit. Typ-
ically, this is the detector which is mounted at the highest or lowest
point at the tank. All other detectors, the Auxiliary Units, are con-
nected to the Main-Unit.
For more information on multi-detector operation please see
page 2–121.
If the following display does not come up, then select it, starting
from the Start menu, via the following path:
ACCESS TO SETUP FULL ACCESS SERVICE
Select PROBE SETTINGS.
Select # OF PROBES.
Enter the number of connected Uni-Probes, e.g. 3.
Up to 8 Uni-Probes can be linked in a network. In our example
we proceed from the assumption that 3 Uni-Probes are linked
with each other, i.e. two Auxiliary Units and one Main Unit.
Save the changes with SEND.
Select
PROBE NO.
Enter 1 (for the first Auxiliary Unit)
(see also page 3–332).
Save the changes with SEND.

Table of Contents