Volume 3 7 Error Handling
3
7.4 Reset
The Uni-Probe can be reset as follows:
1. Software reset
Via
FACTORY SETTINGS on the SERVICE menu.
All parameters will be reset to their defaults. Power the HART
®
Communicator off and on again after using this function.
2. Hardware reset
By pushing the Reset button on the digital board.
The housing must be open! See also Volume 2 of the User’s
Manual, section 3 on page 2–183.
The parameters remain unchanged.
3. Power supply
Turn the power supply off/on to reset the Uni-Probe.
The parameters remain unchanged.
7.5 Operation Modes during Measurement
Operating
mode
Measurement Current output
Normal RUN (normal averaging) Normal
Halt Frozen. The measurement
is stopped by:
–external input HALT
exceeding the count
rate limits in the
parameter Upper Limit
or Lower Limit
Holds the last value
Stop Halted (aborted due to
error message)
Fault current
TEST Display according to test
count rate
Test, follows the test count
rate.
READ-IN Arithmetic averaging of
calibration count rate
Holds the last value