7 Error Handling Volume 3
14 Error Hardware Reset Error in the electronics
Will only be entered in the error
log!
If the error occurs repeatedly, you
have to replace the electronics mod-
ule.
17 Error RTCbackup / Battery Hardware timer does not have the
current date!
If the device has been without
power supply for a longer period
of time, then the buffer capacitor
may be empty.
Check date and update it, if neces-
sary. Supply device with power for 24
hours to charge the capacitor again.
If the error message occurred
while the Uni-Probe has been in
operation for several hours
already, then the hardware timer
is faulty.
You can continue working with the
device anyway, as long as it will be
continuously supplied with power.
Otherwise, please keep in mind that
no decay compensation is carried out
as long as the Uni-Probe is without
power supply. As soon as possible,
you have to replace the digital board
to ensure that the decay compensa-
tion will also work when the device is
switched off.
18 Error RTC Hardware timer faulty Replacing the digital board
19 Error RS-485 Communica-
tion
Communication not possible with
indicated Auxiliary Unit.
Check wiring. If necessary, replace
Main Unit and/or Auxiliary Unit
20 Warning cps > Min Count rate below adjusted mini-
mum
Check input value for minimum count
rate. If necessary, the function can
be disabled by entering 0.
21 Warning cps > Max Count rate above adjusted maxi-
mum
Check input value for maximum
count rate. If necessary, the function
can be disabled by entering 0.
22 Warning Rad. Interference Interference radiation detected,
possibly due to welding check.
Wait for set waiting time. The system
automatically switches to the Run
mode.
23 Fault Current Loop Error at the current output.
The actual value of the current
value does not correspond to the
target value.
Replacing the digital board
Error
number
Error message Probable cause Potential solution